Service Manual
Chapter 5
5‑11
5.5 Machine Offset Screen
1.
Click M.Function in the enlarged maintenance sub-menu.
So, the Manual Operation screen appears.
2.
When clicking Offset button in the Manual Operation screen, the
following screen is shown.
Machine Offset screen
Notice : Be sure to input the machine offset value and Direction,
when setting up or exchanging the personal computer.
(Input the parameter values written on the back of the front panel.)
- Upon inputting the parameter value, click Update button.
- When the setting is completed, click Exit button to go back to the main control
screen.