5.6.4 SP (Sample Pump)
1. Explanation of Function
2. INIT
The washing side of pump : To move to the home position
(the upper limit).
3. PULL(1000uL)
The washing side of pump : Aspiration movement (1000uL).
4. PUSH(1000uL) The washing side of pump : Dispensation movement (1000uL).
5. Move Speed
The washing side of pump : * Move speed pulse rate
6. EV1-ON
Electromagnetic valve EV-1 ON
Line switch to the probe side.
7. EV1-OFF
Electromagnetic valve EV-1 OFF
Line switch to the water tank.
8. INIT
The sample side of pump : to move to the home position
(the upper limit).
9. PULL(15uL)
The sample side of pump : Aspiration movement (15uL)
10. PUSH(15uL)
The sample side of pump : Dispensation movement (15uL)
11. Volume Set
The sample side of pump : Setting of the amount of sample for
aspiration / dispensation.
CW means Dispensation, and CCW means Aspiration.
12. Move Speed
The sample side of pump : * Move speed pulse rate
* shows that setting is not needed.
Manual Function (SP)