Connecting an AS40 Automated Sampler (Optional)
Doc. 031857-03 4/06
B-111
B.19 Connecting an AS40 Automated Sampler (Optional)
1. Place the AS40 to the left
of the ICS-2000 on the
workbench.
2. Open the front door of the
ICS-2000 and thread the
outlet line from the AS40
through the left side slot
at the front of the ICS-
2000.
3. Connect the outlet line
from the AS40 to port
S (5)
of the injection valve
Figure B-108. ICS-2000 Injection Valve
4. Locate the relay control cable (P/N 047946) supplied with the AS40.
5. Remove the 2-position connector plug from the pair of wires labeled
LOAD
on
the relay control cable.
6. Locate the 12-position connector
plug (P/N 923686) in the ICS-
2000 Ship Kit (P/N 064375) (see
7. Attach the red wire from the
relay control cable to position 1
on the plug, and attach the green
wire to position 2. This will be
used to connect the AS40
LOAD
relay to
RELAY OUT 1
on the
ICS-2000.
To attach a wire to the plug, strip the end of the wire, insert it into the plug,
and use a screwdriver to tighten the locking screw.
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.
AS40
Sample
Outlet Line
Figure B-109. 12-Position Connector Plug
Position 1
Position 12
Locking
Screws
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...