5 • Service
Doc. 031857-03 4/06
5-25
5.9
Replacing the Conductivity Cell
1. Turn off the ICS-2000 power.
2. Open the front door and disconnect the tubing from the
CELL IN
and
CELL
OUT
, View A). Loosen the two screws on the cell
front cover. The screws remain attached (see
, View B).
Figure 5-16. Removing the Cell Front Cover
3. Grip the two loosened screws
and pull the cell cover toward
you to remove it from the
component panel.
Do not pull
from the
CELL IN
and
CELL
OUT
fittings.
4. Continue pulling on the cover
until the attached cell assembly
slides out the front of the ICS-
2000 (see
Figure 5-17. Removing the Cell Assembly
Disconnect
Unscrew
Unscrew
View A
View B