Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-71
Filling the EGC CO
3
Mixer with Deionized Water
1. Verify that the EGC-1 and EGC-2 currents are off.
2. Disconnect the EluGen cartridge inlet line (labeled
TO
PUMP/DAMPER
) from the coupler that connects it to the green pump
Figure B-70. EluGen Cartridge Inlet and Pump Outlet
TO PUMP/
DAMPER
(disconnect
this line)
Pump Outlet
(green
tubing)