ICS-2000 Ion Chromatography System
5-86
Doc. 031857-03 4/06
4. Thread the three lines from the EPM (red, yellow, and blue labels)
through the opening on the side of the new holder (see
5. Turn the new EGC holder
upside down (see
NOTE To help retain the
fittings and ferrules
on the lines, remove
the coupler from each
line just before
connecting the line to
a port.
Figure 5-76. EGC Holder (upside down)
6. Connect the lines from the EPM (red, yellow, and blue labels) to the
corresponding lines in the holder.
7. Disconnect the line labeled
TO CR-TC/EPM REGEN IN
(orange label)
from the
REGEN IN
port on the EPM and connect the new
TO CR-
TC/EPM REGEN IN
line (orange label) to the
port.
Figure 5-75. TO CR-TC/EPM Lines Outside of Holder
TO CR-TC/EPM
ELUENT OUT
(yellow label)
TO CR-TC/EPM
ELUENT IN
(red label)
TO CR-TC/EPM
REGEN OUT
(blue label)