ICS-2000 Ion Chromatography System
4-8
Doc. 031857-03 4/06
To troubleshoot:
1. Locate the source of the leak by visually inspecting the tubing, fittings,
and components on the component panel.
2. Tighten fittings or replace tubing and fittings as required. Refer to
for detailed troubleshooting of various types of leaks.
3. After fixing the leak, dry the drip tray thoroughly to prevent the leak
sensor from triggering additional error messages.
If the injection valve fails to switch position within 1 second of being toggled,
the ICS-2000 Moduleware reports an error to Chromeleon and the
“Load/Inject valve error” message appears.
To troubleshoot:
1. If a sequence is being executed, terminate the sequence by selecting
Batch>Stop
on the Chromeleon menu.
2. Turn off the ICS-2000 power briefly and then restart.
3. Try to toggle the valve from Load to Inject by clicking the
Load
and
Inject
buttons on the ICS-2000 Control panel in Chromeleon or on the
touch screen
HOME
page.
4. If the problem persists, contact Dionex for assistance.
This error occurs if a command is issued to control an option that is not
installed.
To troubleshoot:
•
Check the Chromeleon Server Configuration program to verify that the
option is enabled.
LOAD/INJECT VALVE ERROR
OPTION NOT INSTALLED
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...