ICS-2000 Ion Chromatography System
B-50
Doc. 031857-03 4/06
1. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT
line from port
P
(
2
)
on the injection valve.
2. Disconnect the
TO SRS/AES REGEN OUT
line from the suppressor.
3. Use a 10-32 to 1/4-28 coupler (P/N 042806) to connect the
TO INJ
VALVE IN - P/ELUENT OUT
line to the
SRS/AES REGEN OUT
line.
4. Verify that the suppressor is off.
5. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2000 touch screen
SUPPRESSOR
). Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.
Figure B-47. CR-TC Plumbing for Hydration
To avoid damaging the suppressor, always turn off the suppressor
before hydrating the CR-TC. The pump flow is on during conditioning,
however, no flow reaches the suppressor.
Elu
e
nt In
R
e
gen In
Re
g
e
n Out
Eluent
Out
CR-TC
EGC
DEGAS
TO DEGAS
ELUENT OUT
TO CR-TC
REGEN OUT
(blue)
TO SRS/AES
REGEN OUT
TO
C
R-T
C
RE
G
EN
IN
(o
ran
ge
)
WASTE,
GAS
SEPARATOR
DEGAS
ELUENT OUT
TO INJ
VALVE IN - P/
ELUENT OUT
TO DEGAS
ELUENT IN
TO CR-TC
ELUENT IN
(red)
TO CR-TC/EPM
ELUENT OUT
(yellow)
10-32 to
1/4-28 Coupler
Deionized water
in from pump
TO EGC
OUT
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...