ICS-2000 Ion Chromatography System
5-80
Doc. 031857-03 4/06
5.21 Replacing the EGC Holder and Degas Assembly
5.21.1 Disconnecting and Removing the EluGen Cartridge
1. Turn off the power to the ICS-2000.
2. Disconnect the following EGC holder liquid lines:
•
Disconnect the line labeled
TO INJ VALVE IN - P/ELUENT OUT
from either port
P
(
2
) on the injection valve, or if an EGC CO
3
Mixer inlet is installed, from the mixer inlet.
•
If a CR-TC is installed, disconnect the line labeled
TO SRS/AES
REGEN OUT
from the
REGEN OUT
port on the suppressor.
•
Disconnect the line labeled
TO PUMP/DAMPER
from the pump
damper.
•
Disconnect the line labeled
WASTE-GAS SEPARATOR
from the
gas separator waste tube assembly installed on the waste
container.
3. Disconnect the EluGen cartridge cable, and the CR-TC or EPM cable
(if installed) from their respective connectors on the ICS-2000 top
cover.
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), oder eine korrodierende Säure (MSA). Tragen Sie daher
beim Umgang mit Kassette eine Schutzbrille und Handschuhe.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...