10. Push the cables into the opening in the component panel.
11. Align the top of the suppressor
with the guide line printed on the
component panel (see
) and align the slots
on the rear of the suppressor
with the tabs on the panel.
12. Press the suppressor onto the
tabs and then slide it down until
it locks into place. Pull slightly
on the center of the suppressor to
verify that it is securely fastened.
Note
: Some suppressors require
more force to secure them onto
the tabs.
13. Connect the two eluent and two
regenerant lines to the new
suppressor.
14. Close the ICS-2000 front door.
Figure 5-20. Suppressor Guidelines and Mounting
Tabs on Component Panel
5.11 Replacing the Column Heater
1. Turn off the ICS-2000 power.
2. Open the front door of the ICS-2000.
3. Follow the steps below to remove the existing column heater:
a.
Unscrew the two thumbscrews on the heater cover (they remain attached
to the cover).
b. Pull the heater cover straight out to remove it.
1
1
2
3
4
1
Suppressor
Mounting Tabs
Suppressor
Cable Connector
Guideline for
AES
Guideline for
SRS
1
2
3
4
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...