5 • Service
Doc. 031857-03 4/06
5-9
10. To recheck the value, wait at least 15 minutes, and then repeat the
calibration.
11. If you are using the Wellness panel, click
Log
to record the new
calibration value in the Audit Trail.
12. If you performed the calibration from the ICS-2000 touch screen,
record the new calibration values in the Chromeleon Wellness
database. To do this, open the Wellness Panel (see
click
Upload
.
5.1.5
Calibrating the Vacuum Degas Assembly
When to Calibrate
•
After every 6 months of use
1. Open the ICS-2000 Wellness Panel in Chromeleon (see
), or open the ICS-2000 touch screen
PUMP DEGAS
page
2. On the Wellness Panel under
Degas Calibration
, select
Calibrate
, or
touch
Calibrate
on the ICS-2000 touch screen. The degas pump runs
for 90 seconds to allow a vacuum to be created. After 90 seconds, the
Figure 5-6. Pump Degas Page