ICS-2000 Ion Chromatography System
4-16
Doc. 031857-03 4/06
NOTE It is especially important to regularly replace end-line
filters when using aqueous eluents, which may
contaminate the filter with bacteria or algae. The
bacterial buildup may not be visible.
•
Blockages in inlet tubing
Kinked or clogged tubing causes the pump to be “starved” for eluent. Replace
the tubing and fittings (see
•
Dirty check valve
Clean or replace the pump check valve (see
•
Liquid leaks at junction between pump head and pump casting
Use the open-end wrench (P/N 014605) provided in the ICS-2000 Ship Kit
(P/N 057905) to tighten the two acorn nuts that attach the pump head to the
pump housing (see
). Tighten the nuts evenly (12 in-lb torque).
If the leak persists, replace the piston seal (see
•
Liquid leaks from the seal wash port
Replace the piston seal (see
•
Scratched pump piston
Check the pump pistons for scratches and replace if necessary (see
4.4
Pump Does Not Start
•
No power (front
Power
LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace if needed (see
•
No communication between ICS-2000 and Chromeleon
Verify that the USB cable is connected correctly. See
for
installation instructions.
•
Pump is turned off
Turn on the pump from either the Chromeleon Control panel or the ICS-2000
touch screen
HOME
page.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...