Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-57
2. Use the precut tubing assemblies included in the ICS-2000 Ship Kit
to extend the length of three lines as described below:
a.
Disconnect the
TO
CR-TC ELUENT IN
line (red label) from the
coupler that connects it to the
TO CR-TC ELUENT OUT
line (yellow
label). Connect a 32 to 10-32 coupler to the end of the
TO
CR-TC
ELUENT IN
line and connect the 30 cm (12 in) length of black
PEEK tubing (red labels) (P/N 064334) to the coupler (see
b. Connect the 35 cm (14 in) length of black PEEK tubing (yellow
labels) (P/N 064340) to the coupler on the
TO CR-TC ELUENT OUT
line (yellow label).
c.
Disconnect the
TO CR-TC REGEN IN
line (orange label) from the
coupler that connects it to the
TO CR-TC REGEN OUT
line (blue
label). Connect
the
28 cm (11 in) length of clear Teflon tubing
(blue labels) (P/N 064337) to the coupler on the
TO CR-TC REGEN
OUT
line (blue label).
Figure B-52. EPM Eluent and Regen Line Extensions
TO CR-TC/EPM
ELUENT OUT
Line (yellow
label)
TO CR-TC/EPM
ELUENT IN
Line (red label)
TO CR-TC/EPM
REGEN OUT
Line (blue label)
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...