ICS-2000 Ion Chromatography System
5-36
Doc. 031857-03 4/06
5.15 Priming the Pump with Isopropyl Alcohol
NOTE Prime the pump heads with isopropyl alcohol only if the
two standard priming procedures described in
are unsuccessful.
1. Connect a 10-mL syringe
(P/N 054578) filled with
isopropyl alcohol (IPA) to
the port in the primary
pump head (see
2. Open the waste valve on
the secondary pump head
(see
turning the knob one-
quarter to one-half turn
counterclockwise.
Figure 5-29. Priming the Pump Heads with IPA
3. Click
Pump On
on the Chromeleon Control panel to turn on the pump motor.
4. Open the priming valve on the primary pump head by turning it one-quarter to
one-half turn counterclockwise.
5. Use the syringe to slowly push alcohol through the pump.
NOTE Be careful not to push any air trapped in the syringe
through the pump. Check the waste line from the
secondary pump head to verify that there are no air
bubbles.
6. Close the priming valve.
Do not overtighten.
Disconnect the syringe from
the priming valve.
7. Let the pump run for several minutes, to purge alcohol from the pump heads,
and then click
Prime
and flush the heads with the desired eluent.
10 mL Syringe
(P/N 054578)
Priming Valve
(open)
Port
Waste Valve
(open)
Primary
Pump Head
Secondary
Pump Head
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...