5 • Service
Doc. 031857-03 4/06
5-55
1. Disconnect the
TO INJ VALVE IN - P/ELUENT OUT
line from the EGC
CO
3
Mixer inlet.
2. Locate the yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil
(P/N 053765) and a 10-32 to 10-32 coupler (P/N 042627) in the ICS-
2000 Ship Kit (P/N 064375).
3. Use the coupler to connect the
TO INJ VALVE IN - P/ELUENT OUT
line
to one end of the backpressure coil (see
4. Direct the free end of the backpressure coil to a temporary waste
container.
Figure 5-51. Plumbing Connections for Conditioning the
EluGen Cartridge with Carbonate Eluent
EGC Cable
Temporary
Waste
Container
Back-
pressure
Coil
TO INJ VALVE
IN - P/
ELUENT OUT