Connecting the Analog Output (Optional)
Doc. 031857-03 4/06
B-113
the ICS-2000 injection valve are included by default in the program file (see
Figure B-111. Example Program for AS40 Control
B.20 Connecting the Analog Output (Optional)
The analog output connector on the ICS-2000 rear panel outputs an analog voltage
signal proportional to the conductivity measured by the cell. The selected data rise
time (see
) filters the output. The analog output can be connected to
an analog-to-digital converter such as an integrator or other recording device.
Refer to the documentation for the device for connection instructions.
Analog output is 0 to 1V, scaled to the selected range (see
For an example setup for ICS-2000 stand-alone operation with an AS or AS50
(USB) autosampler, see
.
B.21 Pressurizing the Eluent Reservoir (Optional)
The ICS-2000 does not require pressurized reservoirs. However, if eluent is
manually degassed or is sensitive to contamination, Dionex recommends
pressurizing the reservoir with helium or nitrogen. The air regulator accessory
bracket and other items needed for pressurizing the eluent reservoir must be
ordered separately (P/N 060054).
Relay 1 turned on to start
the AS40 Load cycle and
Relay 1 turned off after
load the sample loop
138 seconds
Injection valve
switched to Inject
Injection valve
switched back
to Load
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...