C • Touch Screen Operation
Doc. 031857-03 4/06
C-17
•
To change the low pressure limit, touch the
LOW
field and enter the
desired limit on the number pad.
To select the pressure unit for all touch screen pages that display pressure
values, touch the pressure unit field and select the desired unit on the list
(
PSI
,
bar
, or
MPa
C.7.2
Setting Degas Operating Parameters
By default, the ICS-2000 monitors the degas pressure reading and turns
the degas pump on and off as required. To change the degas operating
option:
1. Select
PUMP
from the menu of pages.
2. Under
Degas Settings
, touch the options drop-down list. A list of
available options appears (see
NOTE Degas options can also be selected from the Chromeleon
Server Configuration program (see
3. Select the desired option:
Figure C-14. Pump Page: Selecting the Pressure Unit
Figure C-15. Pump Page: Selecting Degas Options
Select the
pressure unit
Select the
degas option