ICS-2000 Ion Chromatography System
C-18
Doc. 031857-03 4/06
Always Off
: The degas pump is always off.
Always On
: The degas pump is always on. This setting is reserved for
test purposes by a Dionex Service Representative.
Duty Cycle
: The degas pump cycles on and off according to the times
specified in the
Duty Cycle
and
Time Off
fields.
Duty Cycle
specifies
for how long the degas pump runs during a cycle;
Time Off
specifies
the time between cycles.
Monitor
: The ICS-2000 monitors the degas pressure reading and
turns the degas pump on and off as required.
C.7.3
Controlling the Eluent Valve
The eluent valve controls flow from the
eluent reservoir. During routine operation,
the valve opens and closes automatically
when the pump flow is turned on and off. However, during eluent line
priming, you can use the controls on the
PUMP
page to manually open the
valve, allowing liquid to flow through the lines when the pump is off. See
for eluent line priming instructions.
Select the Degas Mode, Always On option only when testing the vacuum
degas assembly. Do not select the Always On option for routine use.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...