Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-43
2. Set the EluGen cartridge in the service area on top of the ICS-2000
and orient it with the cable toward the EGC holder (see
3. Pull the coiled black
EGC IN
and
EGC OUT
lines out slightly from the
cartridge holder.
4. Disconnect the
EGC
IN
line from the
union that connects
it to the
EGC OUT
line.
5. Connect
EGC IN
to
the EGC
INLET
port
Figure B-39. EGC Inlet Port Connections
Figure B-38. EluGen Cartridge in Service Area
EGC Cable
EGC
IN
INLET
Port