5 • Service
Doc. 031857-03 4/06
5-43
NOTE The backpressure coil connection is temporary; it is
used only during the EluGen cartridge conditioning
procedure described in
.
16. Turn the EluGen cartridge over, with the fittings facing down.
Shake
the EluGen cartridge vigorously and tap it with the palm of your
hand 10 to 15 times. Check that all bubbles trapped in the eluent
generation chamber are dislodged.
17. Slide the EluGen cartridge down
into the holder until it stops.
Make sure the cable and
backpressure coil remain out of
the holder (see
NOTE To avoid crimping the
lines inside the holder, do
not twist the cartridge
before you slide it into
the holder. The lines are
designed to coil inside the
holder.
Figure 5-38. Installing the EluGen Cartridge in the Holder
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), oder eine korrodierende Säure (MSA). Tragen Sie daher
beim Umgang mit Kassette eine Schutzbrille und Handschuhe.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...