ICS-2000 Ion Chromatography System
5-6
Doc. 031857-03 4/06
2. On the Wellness Panel under
Electric Conductivity Cell
Calibration
, click
Offset Cal
, or touch
Run Offset Cal
on the ICS-
2000 touch screen.
3. When the offset calibration is complete, click
Slope Cal
on the
Wellness Panel, or touch
Run Slope Cal
on the ICS-2000 touch
screen.
4. If you are using the Wellness Panel, click
Log
to record the new
calibration values in the Audit Trail.
5. Disconnect the pump output line from the injection valve.
6. Disconnect the line from the suppressor
ELUENT OUT
port to the cell
inlet and connect the pump output line directly to the cell inlet.
7. Verify that the backpressure is at least 7 MPa (1000 psi).
8. If you are using the touch screens to perform the calibration, open the
CONDUCTIVITY CELL
page (see
9. On the Wellness Panel under
Conductivity Cell Calibration
, click
Cell 35 C
, or touch
35 C
on the ICS-2000 touch screen. Allow the cell
to reach this temperature and then wait an additional 5 minutes to let
it stabilize. When the temperature has stabilized, “=” is displayed to
the right of the temperature reading on the
CONDUCTIVITY CELL
page.
Figure 5-4. Conductivity Cell Page
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...