4 • Troubleshooting
Doc. 031857-03 4/06
4-3
4.1
Troubleshooting Error Messages
This error occurs when the column heater temperature is higher than the
maximum allowed. This error may occur if the ICS-2000 is operating in an
extreme environment (greater than 40
°C (104 °F))
.
To troubleshoot:
Refer to
for environmental specifications.
This error occurs when the column heater is unplugged from the component
panel.
To troubleshoot:
1. Check that the column heater is plugged into the component panel.
2. If the error persists, the column heater may be faulty. Replace the column
Suppressor over current
Suppressor over voltage
Suppressor over power
Suppressor not connected
Suppressor stopped for flow rate
COLUMN HEATER EXCEEDS SAFE TEMPERATURE
COLUMN HEATER OPEN CIRCUIT
Alarms and Error Conditions
See
Table 4-4. ICS-2000 Alarms and Error Messages Summary (Continued)