2 • Features
Doc. 031857-03 4/06
2-19
Detector Controls
Autozero
Sets the conductivity to zero.
Acq. Ready
Indicates whether the detector is ready to begin data acquisition.
Total Cond
Displays the total conductivity (the conductivity including the offset
conductivity). To determine the offset, allow the system to equilibrate
after startup. At equilibration, the total conductivity is the background
conductivity (the eluent conductivity before sample injection). Touch
Autozero
to zero the conductivity and offset this background reading
from the total reading.
Signal
Displays the offset conductivity (the total conductivity with the
background conductivity subtracted).
Data Rate
Sets the rate at which data points are recorded.
Cell Temp
Sets the temperature of the cell.
Supp. Type
Selects the type of suppressor installed.
Supp. Current
Sets the current supplied to the suppressor.
Supp. Mode
Turns the suppressor on or off.