ICS-2000 Ion Chromatography System
2-36
Doc. 031857-03 4/06
You can select the concentration of eluent to be generated from either the
Chromeleon Control panel or from the touch screen
HOME
page (see
The allowable eluent concentration depends on several factors: the flow
rate, suppressor type, EluGen cartridge type, and cartridge configuration.
For details, refer to
.
Single-Cartridge Configuration
In the single-cartridge configuration, the ICS-2000 contains one EluGen
cartridge.
Linked Dual-Cartridge Configuration
In the linked dual-cartridge configuration, the ICS-2000 contains two
EluGen cartridges (or one EluGen cartridge and an EPM) that perform as
one cartridge. Note that the allowable eluent concentration for a linked
cartridge is less than when the cartridge is defined as independent. For
EluGen Cartridge
Eluent Concentration Range
K
2
CO
3
0.1 to 15 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow in mL/min
KOH
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
LiOH
0.1 to 80 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 80/flow in mL/min
MSA
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
NaOH
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow mL/min
Table 2-1. Eluent Concentration Ranges for Single-Cartridge
Configurations
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...