ICS-2000 Ion Chromatography System
2-22
Doc. 031857-03 4/06
controls for each cartridge in the dual panel are the same as described for
the single cartridge panel controls (see
Target Concentration
Sets the concentration of eluent generated. If a K
2
CO
3
EluGen cartridge
and an EPM Electrolytic pH Modifier are installed (connected to
EGC-1
and
EGC-2
, respectively), set the
EGC-1 Target Concentration
to the
concentration of K
2
CO
3
required for your application and set the
EGC-
2 Target Concentration
to the concentration of KHCO
3
required.
Applied Concentration
Displays the concentration of eluent currently being generated. For
carbonate/bicarbonate eluent, the
Applied Concentration
for the
K
2
CO
3
cartridge is equal to the total of the target concentrations from
both cartridges. This is the actual concentration of K
2
CO
3
that the
cartridge must generate in order to achieve the desired K
2
CO
3
/KHCO
3
eluent mixture:
Applied Concentration K
2
CO
3
= Target Concentration K
2
CO
3
+ Target
Concentration EPM
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...