ICS-2000 Ion Chromatography System
C-26
Doc. 031857-03 4/06
C.10.3 Setting Analog Out Options
The analog output connector on the ICS-2000 rear panel outputs an
analog voltage signal proportional to the conductivity measured by the
cell. The output is filtered using the selected data rise time (see
). The analog output can be connected to an analog-to-
digital converter device, such as an integrator or a chart recorder.
The
Analog Out
options on the
DETECTOR
you configure the output for your application.
•
Output
: Select
Normal
to output a signal corresponding to the offset
conductivity reading from the detector (see
) and the selected
Analog Out
parameters. Use the other
two settings to calibrate an analog-to-digital converter device. Select
Zero
to set the output signal to zero volts. Select
Full Scale
to set the
output signal to the full-scale voltage (1000 mV).
•
% Offset
: To adjust the zero position of the analog output when it is
plotted, enter a percentage in the
% Offset
field. The value entered is
a percentage of the full-scale analog output. An offset allows a
recording device to plot the signal if it becomes negative. The offset
percentage does not affect the magnitude of the output signal.
•
Polarity
: Select the polarity of the analog output signal: positive (the
default settings) or negative. In applications in which the analyte
output is lower than the background conductance, the polarity must
be negative to have peaks instead of dips on the chromatogram.
•
Range
: Select the range in microSiemens (
µ
S) of a full-scale detector
response. For the ICS-2000, the full-scale voltage is 1000 mV. For
example, if the range is 100
µ
S, then 1
µ
S is equal to 10 mV from the
analog output. The range to use depends on the conductivity readings
expected for the application.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...