ICS-2000 Ion Chromatography System
B-112
Doc. 031857-03 4/06
NOTE If
RELAY OUT 1
is not available, you can connect to
RELAY OUT 2
or to one of the TTL outputs. For
RELAY
OUT 2
, connect the red wire to position 3 and the green
wire to position 4. For
TTL OUT 1
, connect the red wire
to position 5 and the green wire to position 11. For
TTL
OUT 2
, connect the red wire to position 6 and the green
wire to position 12.
8. Connect the relay control cable's 10-position connector plug to the
RELAY
CONTROL
connector on the AS40 rear panel (see
9. Plug the 12-position connector plug into the TTL and relay connector on the
ICS-2000 rear panel (see
). Make sure the red wire connects to
position 1 and the green wire connects to position 2.
The AS40 Load relay is now dedicated to the ICS-2000
RELAY OUT 1
. Use this
relay for running with Chromeleon. When you set up the ICS-2000 application
(see
), commands for controlling the AS40 Load relay and switching
Figure B-110. Example AS40 Autosampler Connection