5 • Service
Doc. 031857-03 4/06
5-63
1. Locate the 10-32 to 1/4-28 coupler (P/N 042806) in the ICS-2000
Ship Kit (P/N 064375).
2. Use the coupler to connect the line labeled
TO INJ VALVE IN -
P/ELUENT OUT
and the line labeled
SRS/AES REGEN OUT
.
3. Verify that the current to the suppressor is turned off.
4. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2000 touch screen
ELUENT GENERATOR
page
). Disabling this option lets you turn on the pump,
without automatically turning on the CR-TC.
5. Turn on the pump and set the flow to the rate recommended for your
application. Pump deionized water through the CR-TC for at least
10 minutes.
6. Turn off the pump and disconnect the coupler.
7. Reconnect the line labeled
TO INJ VALVE IN - P/ELUENT OUT
to
port (
P
)
2
on the injection valve.
8. Reconnect the line labeled
TO SRS/AES REGEN OUT
to the
REGEN
OUT
port on the suppressor.
9. Reconnect the CR-TC and EluGen cartridge cables.
10. If you disabled the
Automatically turn on with pump
option in
, select the option.
To avoid damaging the suppressor, always turn off the suppressor
before hydrating the CR-TC. The pump flow is on during hydrating,
however, no flow reaches the suppressor.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...