D • TTL and Relay Control
Doc. 031857-03 4/06
D-7
D.3
Example Setup for Stand-Alone Operation
This section describes an example setup for
stand-alone operation
(operation
without Chromeleon software) with an AS or AS50 (USB) autosampler and a
third-party analog-to-digital converter and controller. If your configuration
includes different devices, use the following steps as guidelines and refer to your
product documentation for specific connection details.
Connect the AS or AS50 (USB) Autosampler
Follow the instructions in
to connect the AS or AS50 (USB)
autosampler to the ICS-2000 injection valve.
Connect the TTLs and Relays
1. Locate two twisted pairs of
wires (P/N 043598) and 12-
position connector plugs
(P/N 923686) (see
) in the ICS-2000
and AS or AS50 (USB) Ship
Kits.
2. Attach the wires to the
connector plug positions
indicated in
.
To attach a wire to a connector plug, strip the ends of the wire, insert into the
desired connector position, and tighten the locking screw with a screwdriver.
3. Plug the connector plugs into their respective TTL/Relay connectors on the
rear panel of each module.
Make sure the red wires connect to active (+) pins on each module and the
black wires connect to ground (-) pins.
When attaching wires to the connector plug, be careful not to allow
stray strands of wire to short to the adjoining position on the
connector.
Figure D-6. 12-Position Connector Plug
Position 1
Position 12
Locking
Screws
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...