ICS-2000 Ion Chromatography System
G-4
Doc. 031857-03 4/06
Packing
The material that fills a chromatographic column; usually a resin or silica-based
material.
Pellicular Resin
A resin with a solid, nonporous core coated with a thin layer of more porous
material. The exchange sites of pellicular ion exchange resins are located only on
the surface layer of the bead. These resins have a low ion-exchange capacity.
Pre-Column
Synonymous with
Guard Column
.
Regenerant
A dilute acid or base that converts ion exchange sites in a MicroMembrane
Suppressor back to the form that suppresses the eluent conductivity.
Resin
See
Ion-Exchange Resin.
Resolution (R)
A measure of the separation between two sample components. This is expressed
as the ratio of the distance between the two peak maxima to the mean value of the
peak width at the baseline.
R = 2(t
2
- t
1
)/(W
2
+ W
1
)
Where: t
1
and t
2
=the retention times of components 1 and 2,
respectively
W
1
and W
1
= the baseline width of peaks 1 and 2,
respectively (measured in the same units as the retention
time)
R is proportional to the square root of efficiency (N). A value of R = 1.5
represents “baseline separation” of the two peaks.
Retention Time
The time from injection to peak maximum; the basis for identification of a species
in chromatographic analysis.
Separator Column
The column used to perform a chromatographic separation; also called an
analytical column.
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...