ICS-2000 Ion Chromatography System
5-30
Doc. 031857-03 4/06
c.
Unscrew the six thumbscrews on the top metal plate (see
remove the plate.
Figure 5-21. ICS-2000 Column Heater Cover Removed
d. Remove the columns from the column heater.
e.
Pull the column heater straight toward you to remove it from the
component panel.
4. Before installing the new column heater, write down the two calibration
values (offset and slope) recorded on the back of the heater.
5. Align the connector on the back of the new column heater (P/N 058060) with
the connector on the component panel and push the column heater onto the
component panel.
6. Remove the heater cover and the metal plate on the new heater.
7. Install the columns in the new heater and replace the top metal plate and
heater cover.
8. Turn on the ICS-2000 power.
Thumbscrews (6)
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...