ICS-2000 Ion Chromatography System
5-62
Doc. 031857-03 4/06
15. Reinstall the EluGen cartridge in the holder.
16. Remove the cap on the luer adapter by twisting.
Save the cap
.
17. Twist the luer lock and
EGC VENT
line into the luer adapter.
18. Do not connect the EluGen cartridge cable until after hydrating the
CR-TC.
Hydrating the CR-TC
The CR-TC must be hydrated before initial use by running ASTM
filtered, Type 1 (18 megohm-cm) deionized water through the CR-TC for
10 minutes, while bypassing the columns and suppressor.
illustrates the liquid flow path through the CR-TC during the
hydration procedure.
Figure 5-57. CR-TC Plumbing for Hydration
Elu
e
nt In
R
e
gen In
Re
g
e
n Out
Eluent
Out
CR-TC
EGC
DEGAS
TO DEGAS
ELUENT OUT
TO CR-TC
REGEN OUT
(blue)
TO SRS/AES
REGEN OUT
TO
C
R-T
C
RE
G
EN
IN
(o
ran
ge
)
WASTE,
GAS
SEPARATOR
DEGAS
ELUENT OUT
TO INJ
VALVE IN - P/
ELUENT OUT
TO DEGAS
ELUENT IN
TO CR-TC
ELUENT IN
(red)
TO CR-TC/EPM
ELUENT OUT
(yellow)
10-32 to
1/4-28 Coupler
Deionized water
in from pump
TO EGC
OUT
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...