Priming the Pump
Doc. 031857-03 4/06
B-31
B.13 Priming the Pump
1. Verify that the eluent reservoir is filled, the reservoir cap is installed and
handtightened, and the liquid line from the ICS-2000 to the reservoir cap is
connected (see
2. Verify that the waste lines are directed to a waste container.
3. The priming procedure consists of two parts:
•
Priming the eluent lines with a syringe (see
procedure at initial installation, after changing eluents, or when eluent
lines are empty.
•
Priming the pump heads with the
Prime
Perform this procedure after the eluent lines are primed.