Precautions
Storage
Maintenance and Inspection Items
Cautions
• Clean the system before and after using. When disinfecting, do not use a sterilizer. Use a soft cloth (damp with
disinfectants) to wipe the system, and wipe off the disinfectants with a soft cloth (damp with cold/lukewarm water), and then
thoroughly wipe off any moisture with a dry soft cloth. Follow the instructions for use of each disinfectant (regarding degree
of dilution, etc.). Examples of disinfectants (ingredient names) which can be used are listed below.
Chlorhexidine gluconate/Benzalkonium chloride
• If this system is tainted with blood, wipe the system immediately with a damp (not wet) soft cloth. Be sure to wear gloves to
prevent infection. [Infection may occur.]
• When liquids (bloods or drug solution, etc.) come into contact with the pump receptacle of the Drive Motor or the Hand
Crank, immediately wipe them off and clean the devices thoroughly. [If coagulated liquids (bloods or drug solution, etc.) are
adhered to the pump receptacle of the Drive Motor or the Hand Crank, the slide hook may not move.]
• Turn the system off and disconnect the AC power cable and cables (the Drive Motor, the ABD/Flow Sensor, the
Temperature Sensor Cable, the Pressure Sensor Cable, the Fast Clamp, the LAN cable, the CDI Communication Cable)
before cleaning. [Otherwise the system may fail or an electrical shock may occur.]
• Do not wash this system with running water or while immersed in water as this may cause damage or failure of the system.
[This system does not have a waterproof structure.]
• Do not wipe with organic solvent such as alcohol or thinner, or with povidone-iodine. [Using organic solvent or any
disinfectants other than those permitted for use may result in damage to or failure of this system.]
• Do not subject this system to ethylene oxide gas (EOG) sterilization or high pressure steam sterilization, or immerse this
system in disinfectant solution. [The system may fail.]
• Do not let anesthetic drugs come into contact with the system or accessories. [The system or accessories may be
damaged.]
• Do not use an electric dryer to dry this system. [This system may be damaged.]
• Do not use any replacement parts other than those specified. [This may result in decreased performance.]
• Check displays to ensure this system will operate properly. [This system may not operate as intended.]
• Do not store the battery in a discharge status. [Storing the battery left in a discharge status may cause degradation,
resulting in making the battery unavailable in emergency situations.]
• Periodically check the deterioration status of the battery. [Depending on the deterioration status of the battery, the capacity
may not be enough for the system to be powered by the battery, even if all of the battery indicator lamps on the control
panel and the battery status indicator on the LCD touch panel display light up when connected to the AC power source
(even if the battery voltage is normal).]
25
• Do not expose this system to direct sunlight or ultraviolet irradiation for a long period of time. [The exterior may experience
colour change, deformation or deterioration.]
• Do not store this system in a place with a high level of vibration, dust, mist or corrosive gas.
• Do not store this system in a place where an environment of atmospheric pressure, temperature, humidity, ventilation, or
corrosive conditions may cause an adverse effect.
• Do not store this system in a storage area for chemicals or in a place where gases are generated.
• Keep magnetic media, such as hard disks, away from the Drive Motor and the Hand Crank as these devices contain
magnets. [Magnetic medium may be damaged.]
• Do not store this system under high temperature and high humidity.
Cautions
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...