Alarms and Troubleshooting
Occurrence
Cause
Action
The temperature is displayed as
“--.-”.
The temperature is not stable.
• The Temperature Sensor Cable is not
properly connected.
• Reconnect the Temperature Sensor Cable.
(Refer to page 37.)
• The CAPIOX Luer Thermistor is not
properly connected.
• Reconnect the CAPIOX Luer Thermistor.
(Refer to page 46.)
• The CAPIOX Luer Thermistor has failed.
• Replace the CAPIOX Luer Thermistor.
• The Temperature Sensor Cable has failed.
• Contact Terumo’s service representative.
The Clamp ready lamp does not
light up.
• The Fast Clamp is not properly connected.
• Connect the Fast Clamp. (Refer to
page 38.)
• The fast clamp function is not on.
• Enable the fast clamp function. (Refer to
page 90.)
• The safety connection which allows the
Fast Clamp to operate has not been set.
• Set the safety connection. (Refer to
page 91.)
• The tube has not been securely set.
• The lid is not properly closed.
• Set the tube and close the lid. (Refer to
page 51.)
The Fast Clamp cannot be
clamped from the button on Fast
clamp screen.
• The tube has not been securely set.
• The lid is not properly closed.
• Set the tube and close the lid. (Refer to
page 51.)
• The fast clamp function is not on.
• Enable the fast clamp function. (Refer to
page 90.)
• The Drive Motor is running.
• Execute the clamp after stopping the Drive
Motor.
The key lock function does not
activate.
• An alarm other than the On Battery Alarm
is issued.
• Resolve the cause of the alarm. (Refer to
page 134.)
• Log data is being exported.
• The key lock function does not activate
while log data is being exported. This is
not a malfunction.
The count-down ended sound
does not sound.
• An alarm is issued.
• When an alarm is issued, the alarm sound
has priority over the count-down ended
sound. This is not a malfunction.
The button that exports history
data to a USB memory drive does
not function.
• The Drive Motor is running.
• Output data to the USB memory drive after
stopping the Drive Motor.
“Failed to output the log to the
USB.” is displayed.
• The USB memory drive cannot be written.
• Enable the USB memory drive writable,
and try to export the data again.
A USB memory drive is not
recognized.
• Removal of the previous USB memory
drive was not properly executed.
• Turn off the power of the Controller, and
turn on the power again.
• Enable the USB memory drive writable,
and try to export the data again.
The password to open admin
setup is lost.
• Contact Terumo’s service representative to
initialize the password.
133
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...