Specifications
Specifications
Code number
CX
*
XSA CX
*
XSA0171 CX
*
XSA0173 CX
*
XSA0251
Sterilization
Ethylene oxide gas
Priming volume
550 mL
620mL 620mL 740mL
Centrifugal pump
SP45
SL45 SL45 SL45
Maximum blood flow rate
7 L/min
Minimum blood flow rate
0.5 L/min
Maximum pressure
1000 mmHg (133 kPa)
Connector for venous cannula
Connector for arterial cannula
Lock connector
Lock connector
Items
Specifications
Oxygenator
Fiber surface area/material
Heat exchanger surface area/material
Purge port
Gas port (Gas in)
Gas port (Gas out)
Water port
Thermistor probe
Approx. 2.5m
2
/ Poly methyl pentene
Approx. 0.22m
2
/ Stainless steel
Luer port
1/4" (6.4 mm)
1/4" (6.4 mm)
1/2" (12.7mm) Couples connect fitting
Blood outlet port
Centrifugal pump
Maximum pump speed
Maximum outlet pressure
Applicable console
Applicable drive motor
3000rpm
800mmHg (107kPa)
CAPIOX SP Pump Console SP-101
(Code number: ME
*
SP101K, ME
*
SP101K2, ME
*
SP101K3,
ME
*
SP101S, ME
*
SP101S2, ME
*
SP101S3)
CAPIOX Centrifugal Pump Controller SP-200
(Code number: ME
*
SP200C, ME
*
SP200C2, ME
*
SP200C3)
CAPIOX SP Pump Drive Motor
(Code number) ME
*
SP101M, ME
*
SP200M
Applicable flow sensor
CAPIOX SP Pump Flow Sensor 3/8" (9.5 mm)
(Code number: ME
*
FS38)
ABD/Flow Sensor
(Code number: ME
*
SPFAS01, ME
*
SPFAS02)
Accessories (Sold separately)
Applicable holder
CAPIOX EBS Circuit Holder (Code number: XX
*
EB04)
Applicable cannula
CAPIOX EBS Cannula KIT
For the code number, refer to the Instructions for Use of the Cannula Kit.
Applicable temperature monitor
YSI temperature YSI400 Series (Measurement Specialties Inc.)
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...