Parts Description
ABD/Flow Sensor
Product name: CAPIOX Centrifugal Pump Controller SP-200 ABD/Flow Sensor
Manufacturer: Transonic Japan
(Catalogue number: ME
*
SPFAS01 (for standard tube circuit), ME
*
SPFAS02 (for soft tube circuit))
An ultrasonic transmission sensor and a receiving sensor are built in. The flow is measured by the ultrasonic transit method, and
air bubbles are detected by the change in the ultrasonic receiving level.
The ABD/Flow Sensor is attached to the blood circuit tube (inner diameter 9.5 mm (3/8 inches)).
Plug
Cable
Clamp-on ABD/Flow sensor
Fast Clamp
(Catalogue number: ME
*
SPCLP01)
When air bubbles are detected or back flow occurs, the blood circuit is instantly blocked off with the plunger.
The Fast Clamp is attached to the blood circuit tube (inner diameter 9.5 mm (3/8 inches)).
Lid
Clamp ready lamp
Plunger
Latch
Cable
Lever
Plug
Fixing block
Temperature Sensor Cable
(Catalogue number: XX
*
SPCBL011 (blue), XX
*
SPCBL012 (red))
When measuring temperature, connect the applicable temperature sensor (CAPIOX Luer Thermistor) to the Controller.
Plug
Temperature sensor
connector
Cable
14
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...