Various menu settings
The following admin setup values can be set:
• The on/off status of the temperature measuring function
• Threshold value of the Pressure Alert (setup range: 0.0 to 50.0°C)
For details about how to set, refer to “Setting Temperature Measuring Function” (page 88).
• The changes to the setting will take effect after restarting.
The admin setup value for the fast clamp function can be set on/off.
For details about how to set, refer to “Setting the Fast Clamp” (page 90).
• The changes to the setting will take effect after restarting.
The following admin setup values can be set:
• The on/off status of the air bubble detector function
• Air bubble detection sensitivity (High/Medium/Low)
For details about how to set the preset value of the air bubble detector function, refer to “Setting the Air Bubble Detector Function”
(page 90).
This section describes the setup procedures for sensitivity of the air bubble detector function.
• The on/off setting for the air bubble detector function will take effect after restarting. The sensitivity setting will take effect
instantly.
1.
Tap the Air Bubble icon. (Figure 3-69)
The Air Bubble screen opens.
Figure 3-69
2.
Tap the Sensitivity you want to set. (Figure 3-70)
[High]:
Air bubbles larger than 0.05 cm
3
are detected (at flow of 4.20 LPM
(L/min) or less)
[Medium]: Air bubbles larger than 0.2 cm
3
are detected (at flow of 7.00 LPM
(L/min) or less)
[Low]:
Air bubbles larger than 0.5 cm
3
are detected (at flow of 7.00 LPM
(L/min) or less)
Figure 3-70
104
Temperature Preset
Note
Setting the Fast Clamp
Note
Setting the Air Bubble Detector Function
Note
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...