Operation Procedure
Weaning
1.
When weaning, operate the clamp while paying attention to
back flow.
2.
Turn the motor speed knob counter-clockwise to change
the motor speed to 0 RMP to stop the Drive Motor.
(Figure 1-63)
For details about how to stop the motor, refer to page 8.
• By performing the above procedure, some alarms and alerts are displayed on the
message area. This, however, is part of the end process and not a malfunction.
Figure 1-63
• The motor speed turns to “0”, and the flow turns to “0.00”.
If using the Fast Clamp
Close the plunger of the Fast Clamp before finishing the case and store it in a closed (up) position.
3.
Tap the setup icon (Figure 1-64), and tap the Fast clamp
icon on the Setup screen. (Figure 1-65)
The Fast clamp screen opens.
Figure 1-64
Figure 1-65
57
• When weaning, be sure to clamp the blood line and then stop the rotation of the centrifugal pump. If the rotation of the
centrifugal pump stops without clamping the blood line, blood may flow back.
• Do not rotate the centrifugal pump for a long period of time (multiple hours) while clamping the blood outlet line of the
centrifugal pump. The priming solution may be denatured due to heating. The blood may be damaged due to heating.
• After circulation, make sure to rotate the motor speed knob to stop the Drive Motor and then turn off the power.
• When the Drive Motor is removed, place a cap on the plug.
Cautions
Note
Check
Motor speed knob
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...