When an alarm/error occurs, the factors below may be the cause. Follow the procedure for dealing with each alarm/error as it
arises. If the abnormality persists, the system may have failed. Stop using the system immediately and contact Terumo’s service
representative.
Troubleshooting
Occurrence
Cause
Action
The power does not turn ON.
• The battery charge level is low and the AC
power source is not connected.
• Connect the AC power source. (Refer to
page 31.)
When the AC power source is cut
off, the system does not switch
to the battery-powered operation
and the power turns off.
• The battery has little power left.
• Charge battery. (Refer to page 72.)
• The battery has deteriorated or is
damaged.
• Contact Terumo’s service representative.
The AC power lamp does not light
up.
(AC power source cannot be
connected)
• The AC power source is not properly
connected.
• Reconnect the AC power cable. (Refer to
page 31.)
• The AC power cable is internally
disconnected.
• Contact Terumo’s service representative.
The charge lamp does not light
up.
(Cannot charge)
• The battery is fully charged.
• When battery is fully charged, it cannot be
charged further. There is no abnormality.
• The battery is deteriorated or in failure.
• Contact Terumo’s service representative.
Even though charging is
completed (the charge lamp is
off), the battery charge level is
less than 100%.
• The battery has deteriorated.
• When battery charge level is less
than 70%, contact Terumo’s service
representative for replacement of the
battery.
• The battery is damaged.
• An abnormality occurred in the system.
• Contact Terumo’s service representative.
The time indication is not correct.
• The time lags due to time accuracy.
• Set the time again. (Refer to page 109.)
The motor speed is displayed as
“----”, and the centrifugal pump
does not rotate.
• The Drive Motor is not properly connected.
• Reconnect the Drive Motor. (Refer to
page 33.)
• The Drive Motor has failed.
• Contact Terumo’s service representative.
Auto-priming does not start.
• The Drive Motor is running.
• Execute auto-priming after stopping the
Drive Motor. (Refer to page 49.)
• The battery charge level is low and the AC
power source is not connected.
• Execute auto-priming after connecting the
AC power source.
The flow is displayed as “-.--”.
The flow does not stabilize.
• The ABD/Flow Sensor is not properly
connected.
• Reconnect the ABD/Flow Sensor. (Refer to
page 36.)
• The ABD/Flow Sensor is not properly
attached on the tube.
• Reattach the ABD/Flow Sensor on the
tube. (Refer to page 44.)
• The detection window of the ABD/Flow
Sensor is dirty.
• Remove the ABD/Flow Sensor from
the tube to clean (Refer to page 84.),
reapply vaseline and reattach on the tube.
(Refer to page 44.)
• The ABD/Flow Sensor has failed.
• Contact Terumo’s service representative.
The pressure is displayed as “---”.
The pressure does not stabilize.
• The Pressure Sensor Cable is not properly
connected.
• Reconnect the Pressure Sensor Cable.
(Refer to page 37.)
• The pressure sensor is not properly
connected.
• Reconnect the pressure sensor. (Refer to
page 46.)
• The pressure sensor has failed.
• Replace the pressure sensor.
• The Pressure Sensor Cable has failed.
• Contact Terumo’s service representative.
Calibration cannot be
implemented.
“0” does not display even after
calibration is implemented.
• Pressure within the usage environment is
not stable.
• Calibrate the system in an environment
where the pressure is stable.
132
Alarms and Troubleshooting
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...