O
2
transfer
CO
2
transfer
Heat exchanger performance factor
Pump performance
SP45 SL45
Data
Blood Temperature:37
±
1
℃
:Bovine Blood
:12
±
1g/dL
:7.4
:65
±
5%
:45
±
5mmHg
:0
±
5mEq/L
:1.0
<Conditions>
Blood
Hb
pH
SvO2
PvCO2
B.E.
V/Q
<Conditions>
Blood
Hb
pH
SvO2
PvCO2
B.E.
V/Q
:Bovine Blood
:12
±
1g/dL
:7.4
:65
±
5%
:45
±
5mmHg
:0
±
5mEq/L
:1.0
Blood Temperature:37
±
1
℃
<Conditions>
Blood
Hb
Tbi
Twi
Water flow rate
:Bovine Blood
:12
±
1g/dL
:30
±
1
℃
:40
±
1
℃
:15 L/min
Blood flow rate(L/min)
8
6
4
2
0
500
400
300
200
100
0
500
400
300
200
100
0
0
2
4
6
8
Blood flow rate(L/min)
O
2
trand
fer rat
e
(mL/min)
CO
2
trab
sfer r
a
te(mL/min)
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...