Alarms and Troubleshooting
Message display
Cause/Occurrence
Action
Priority
Error code
Clamp Occlusion
Error*
3
Detected a clamp operation
abnormality of the Fast
Clamp.
a
Check that the tube is properly placed
and that the lid is properly closed.
b
Clamp the tube using forceps when
necessary.
* When cancelling the alarm, disable the
fast clamp function.
High
E901
System Error
Detected an equipment
abnormality requiring
inspection after use.
The system may have failed. Do not
disconnect the AC power source. In
addition, contact Terumo’s service
representative as soon as the current case
ends.
Medium
Battery-powered
operation started
Detected that battery-
powered operation has
started.
If this alarm is issued unintentionally, check
the supply of AC power.
Medium
When intentionally stopping the supply of
AC power due to transportation or other
purposes, cancel the Battery-powered
operation started Alarm. (Tap [Battery-
powered operation started]
Confirmation
message displays
Tap [OK])
T1:Low Temperature
Alert*
2
*
3
Detected that the
temperature has fallen below
the setup temperature alert
threshold.
a
Check that there is no abnormality in
the blood circuit.
b
Check the hot and cold water tank.
Medium
T2:Low Temperature
Alert*
2
*
3
T1:High Temperature
Alert*
2
*
3
Detected that the
temperature has exceeded
the setup temperature alert
threshold.
T2:High Temperature
Alert*
2
*
3
On Battery
Detected that the operation
is continuously battery-
powered after confirming
the operation of the Battery-
powered operation started
Alarm.
Restart the AC power supply as soon as
possible while monitoring the battery charge
level.
Low
T1:Temp. Sensor
Disconnected*
2
*
3
Detected that the
temperature sensor (the
CAPIOX Luer Thermistor) is
disconnected.
Connect the Temperature Sensor Cable
to the temperature sensor (the CAPIOX
Luer Thermistor). Connect the cable to the
Controller.
Low
Reconnect the Temperature Sensor Cable
to the temperature sensor (the CAPIOX
Luer Thermistor). Connect the cable to the
Controller.
T2:Temp. Sensor
Disconnected*
2
*
3
Replace the Temperature Sensor Cable or
the temperature sensor (the CAPIOX Luer
Thermistor).
P1:Pressure Sensor
Disconnected*
2
*
3
Detected that the pressure
sensor is disconnected.
Connect the Pressure Sensor Cable to the
pressure sensor. Connect the cable to the
Controller.
Low
Reconnect the Pressure Sensor Cable to
the pressure sensor. Connect the cable to
the Controller.
P2:Pressure Sensor
Disconnected*
2
*
3
Replace the Pressure Sensor Cable or the
pressure sensor.
*
2
: An alarm is not issued until the motor speed indicated value exceeds the coast speed after the Drive Motor starts running.
*
3
: An alarm is issued only when the associated function is set to [On].
136
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...