Various menu settings
The on/off status of the fast clamp function can be set.
1.
Tap the Fast clamp icon. (Figure 3-27)
The Fast clamp screen opens.
Figure 3-27
2.
After tapping [On] or [Off], tap [OK] to complete the setting.
(Figure 3-28)
Figure 3-28
• If you do not tap [OK], the changes to the setting will not take effect.
<Indication of the clamp status icon>
The icon indication changes in accordance with the status of the Fast Clamp.
For details, refer to “LCD Touch Panel Display (When Operating Normally)” (page 9).
The on/off status of the air bubble detector function can be set.
1.
Tap the Air Bubble icon. (Figure 3-29)
The Air Bubble screen opens.
Figure 3-29
90
Setting the Fast Clamp
Note
Setting the Air Bubble Detector Function