For Medical Staff
Table 5 – Recommended separation distances between portable and mobile RF communications equipment and
the ME EQUIPMENT or ME SYSTEMS – for LIFE-SUPPORTING ME EQUIPMENT and ME SYSTEMS (see 5.2.2.2)
Recommended separation distances between portable and mobile RF communications equipment and
the CAPIOX Centrifugal Pump Controller SP-200
The CAPIOX Centrifugal Pump Controller SP-200 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the CAPIOX Centrifugal Pump Controller SP-200 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the CAPIOX Centrifugal Pump Controller SP-200 as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
outside ISM bands
d=1.2
√
P
150 kHz to 80 MHz in
ISM bands
d=1.2
√
P
80 MHz to 800 MHz
d=1.2
√
P
800 MHz to 2.5 GHz
d=2.3
√
P
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
Notes
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz
to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
3. An additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation
distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz
to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is
inadvertently brought into patient areas.
4. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
144
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...