Other Operation Procedures
Refer to the table below for an explanation of each item.
Item
Overview
Description
Remark
Data start
Indicates the data start.
Year/month/
day/hour/
minute/second,
sequence
number, MAC
address
AD four digits for year, 01 to 12 for month, 01 to 31 for day,
00 to 23: 24-hour clock for hours, 00 to 59 for minutes,
00 to 59 for seconds, and a built-in clock of the Controller is
used as a calendar.
The sequence number is cycled by taking numerical numbers
between 0000000000 to 4294967295, and 0 suppression is
not performed.
MAC addresses are characteristic numbers of the Controller.
Flow value
Indicates flow values.
Flow value,
unit
The flow value is from -9.99 to 9.99 or NG, the last number
0 is not omitted, and when the flow is NG, an invalid value is
indicated.
The unit is LPM (L/min).
Pressure 1 ch
Indicates pressure value
1 ch.
Pressure
value, unit
The pressure value is from -250 to 900 (mmHg), from -33.3
to 120.0 (kPa), or NG, and the value is zero-suppressed, and
when NG, an invalid value is indicated.
The unit is mmHg or kPa.
Pressure 2 ch
Indicates pressure value
2 ch.
Temperature
1 ch
Indicates temperature value
1 ch.
Temperature
value, unit
The temperature value is from 0.0 to 50.0 or NG, the value is
zero-suppressed, the last number 0 is not omitted, and when
the temperature value is NG, an invalid value is indicated.
The unit is deg C.
Temperature
2 ch
Indicates temperature value
2 ch.
Motor speed
Indicates motor speed.
Motor speed,
unit
The motor speed is from -9999 to 9999 or NG, the value is
zero-suppressed, and when the motor speed is NG, an invalid
value is indicated.
Alarm
Indicates alarms.
Alarm status
Indicates an alarm occurrence status of 192 bits.
Data end
Indicates the data end.
80
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...