Maintenance and Inspections
5.
Unplug the AC power cable and change the motor speed
to 3000 RPM with the centrifugal pump not attached to the
Drive Motor. At this time, the Battery-powered operation
started Alarm is issued. Tap [Battery-powered operation
started] displayed on the message area and tap [OK] after
confirming the message displayed. (Figure 5-14)
Measure the elapsed time with a stopwatch.
6.
Maintain the status in step 5 until the power of the
Controller turns off.
Figure 5-14
• The system can operate for more than three hours.
7.
Plug in the AC power cable to charge the battery until the
charge lamp turns off.
Measure the elapsed time with a stopwatch.
• The charge lamp turns off within six hours.
• Since the battery is built-in, contact Terumo’s service representative when replacing it.
128
Check
Check
• Do not store the battery in a discharge status. Storing the battery left in a discharge status may cause degradation,
resulting in making the battery unavailable in emergency situations.
• Do not disassemble, make alterations (including actions that interfere with the functionality or performance such as taping
the LCD or a movable part) to, or repair this system. This may result in failure, damage or device performance degradation
of this system.
Cautions
Note
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...