Other Operation Procedures
2.
Clamp the blood line of the patient side of the Fast Clamp
with forceps. (Figure 2-9)
Figure 2-9
3.
Pull up the lever of the Fast Clamp until you hear a
click, and remove the occlusion with the Fast Clamp.
(Figure 2-10)
The plunger is set at the open position, and “Back Flow: Clamp Close” on the
message area is turned off.
• After pulling up the lever, the lever has returned to the original position.
• If the lever does not return to the original position, pull down the lever. If it does
not return to the original position, a failure may have occurred.
Stop using the system immediately, and contact Terumo’s service representative.
Figure 2-10
4.
Keep clamping the blood line with forceps. Turn the motor
speed knob to adjust the motor speed to a speed that does
not cause back flow. Gradually release the clamp of the
blood line.
The status lamp is lit in green, and the alarm sound stops.
“Back Flow: Pump Stop” on the message area is turned off.
5.
While monitoring the flow, adjust the motor speed to
achieve the necessary flow.
71
Check
Note
Blood line
Fast Clamp
Forceps
Lever
S
ide of
th
e
patient
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...