System Specifications (Controller)
Product name
CAPIOX Centrifugal Pump Controller SP-200
Catalogue number
ME
*
SP200C (Identification code: SP200C
*
): Controller
ME
*
SP200C2 (Identification code: SP200C2
*
): Controller
ME
*
SP200C3 (Identification code: SP200C3
*
): Controller
ME
*
SP200M: Drive Motor
ME
*
SPFAS01: ABD/Flow Sensor (for standard tube circuit)
ME
*
SPFAS02: ABD/Flow Sensor (for soft tube circuit)
ME
*
SPCLP01: Fast Clamp
XX
*
SPCBL011: Temperature Sensor Cable (blue)
XX
*
SPCBL012: Temperature Sensor Cable (red)
XX
*
SPCBL021: Pressure Sensor Cable
XX
*
SPCBL031: CDI Communication Cable
XX
*
SPCLP01: Fast Clamp Attachment (for EBS)
XX
*
SPCLP02: Fast Clamp Arm
XX
*
SPCRT01: Cart
Drive Motor drive function
Motor speed range: 0 - 3000 RPM
Motor speed error: ±1.5% (at 1000 RPM or more)
Flow measuring function
Measuring range:
-9.99 to 9.99 LPM (L/min)
Measuring accuracy: ±10% (from -7.00 to -1.01 LPM (L/min) and 1.01 to
7.00 LPM (L/min))
±0.1 LPM (L/min) (-1.00 to 1.00 LPM (L/min))
When attaching to dedicated tube (inner diameter 9.5 mm (3/8 inches)),
and when hematocrit value is 40%, liquid temperature is 37°C and
ambient temperature is 25°C
Air bubble detector function
Low sensitivity:
Detects air bubbles of 0.5 cm
3
or more (at flow
7.00 LPM (L/min) or less)
Medium sensitivity: Detects air bubbles of 0.2 cm
3
or more (at flow
7.00 LPM (L/min) or less)
High sensitivity:
Detects air bubbles of 0.05 cm
3
or more (at flow
4.20 LPM (L/min) or less)
Detection sensitivity can be turned on/off and set
Temperature measuring function (2 channels)
Measuring range:
0.0 - 50.0°C
Apparatus accuracy: ±0.3°C
Function can be turned on/off and set for each channel
Pressure measuring function (2 channels)
Measuring range:
-250 to 900 mmHg ( -33.3 to 120.0 kPa)
Apparatus accuracy: ±3% (-250 to -101 mmHg and 101 to 900 mmHg)
±3 mmHg (±0.4 kPa) (-100 to 100 mmHg)
Function can be turned on/off and set for each channel
Auto-priming function
Run the Drive Motor intermittently to eliminate air bubbles. The
available operation settings are as follows:
Motor speed: From 1000 to 3000 RPM
Motor on time: From 0.5 to 10.0 seconds
Motor off time: From 0.5 to 10.0 seconds
Blood circuit occlusion function
This function occludes the blood circuit by using the Fast Clamp when
air bubbles or back flow is detected.
Function can be turned on/off
138
Specifications
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...