Alarms and Troubleshooting
Message display
Cause/Occurrence
Action
Priority
Error code
Low Flow Alarm*
2
Detected that flow has fallen
below the setup flow alarm
threshold.
a
Check that there is no abnormality in
the blood circuit.
b
Adjust the motor speed to render
appropriate flow.
High
High Flow Alarm*
2
Detected that flow has
exceeded the setup flow
alarm threshold.
Air Bubble Detected
Alarm*
2
*
3
Detected air bubbles that
correspond to the setup
sensitivity.
a
Check that the ABD/Flow Sensor is not
detached from the tube.
b
Cancel the Air Bubble Detected Alarm.
(Tap [Air Bubble Detected Alarm]
Confirmation message displays
Tap
[OK] after confirmation)
High
a
Remove air bubbles.
b
Cancel the Air Bubble Detected Alarm.
(Tap [Air Bubble Detected Alarm]
Confirmation message displays
Tap
[OK] after confirmation)
Air Bubble: Clamp
Close
Detected that the Fast
Clamp operated as the
safety connection of the Air
Bubble Detected Alarm and
circulation stopped.
Handle according to “
c
When the safety
connection of the Air Bubble Detected
Alarm is set to “Clamp+Pump stop””
(page 68).
High
Air Bubble: Pump
Stop
Air Bubble: Pump
Coasting
Detected that the safety
connection of the Air Bubble
Detected Alarm operated and
the motor speed changed to
the coast speed.
Handle according to “
b
When the safety
connection of the Air Bubble Detected
Alarm is set to “Pump Coast”” (page 67).
High
P1:Low Pressure
Alarm*
2
*
3
Detected that the pressure
has fallen below the setup
pressure alarm threshold.
a
Check that there is no abnormality in
the blood circuit.
b
Adjust the motor speed while checking
the flow.
High
P2:Low Pressure
Alarm*
2
*
3
P1:Over Pressure
Alarm*
2
*
3
Detected that the pressure
has exceeded the setup
pressure alarm threshold.
P2:Over Pressure
Alarm*
2
*
3
P1 Low: Pump
Coasting
Detected that the safety
connection of the Pressure
Alarm operated and the
motor speed changed to the
coast speed.
Handle according to “
a
When the safety
connection of the Pressure Alarm (over
pressure/low pressure) is set to “Pump
Coast”” (page 66).
High
P2 Low: Pump
Coasting
P1 Over: Pump
Coasting
P2 Over: Pump
Coasting
Clamp Closed*
3
Detected that the blood
circuit has been clamped
by the Fast Clamp through
an operation other than the
safety connection of the Air
Bubble Detected Alarm or
the Back Flow Alarm.
Operate the lever of the Fast Clamp and set
the plunger to the release position. (Refer to
“Attaching the Fast Clamp” (page 51).)
High
E984
*
2
: An alarm is not issued until the motor speed indicated value exceeds the coast speed after the Drive Motor starts running.
*
3
: An alarm is issued only when the associated function is set to [On].
135
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...