Precautions
Cautions
• Since this system does not have an airtight structure, it should not be used or stored in an active gas environment
(including sterilizer gas), nebulizer-sprayed environment, high-humidity environment, etc. [Taking the system to these
environments leads to the electronic components inside the system being affected and there may be subsequent damage
and time degradation which will cause failure of this system.]
• This system should not be used in a flammable environment. [The system may catch fire or explode.]
• Do not use this system as a suction device. [The system equipment is not intended to perform sucking operations.]
• Do not disassemble, make alterations (including actions that interfere with the functionality or performance such as taping
the LCD or a movable part) to, or repair this system. [This may result in failure, damage or device performance degradation
of this system.]
• Execute maintenance and inspections periodically for devices in the backup kit. When executing maintenance and
inspections, use the Drive Motor and the AC power cable included in the backup kit. [Due to a low frequency of use, the
abnormality may not be detected.]
• Prevent physical shock to the pump receptacles of the Drive Motor and the Hand Crank. [The internal magnets of these
devices may be damaged.]
• If there is physical damage, such as a crack, to the centrifugal pump and the detection window of the ABD/Flow Sensor, do
not use the device. [Leakage of blood may occur, and decreased performance may result.]
• When attaching or detaching the Drive Motor or the ABD/Flow Sensor to or from the Controller, turn off the power. [It may
cause a failure for the Controller.]
• When attaching or detaching the Drive Motor to or from the Backup Controller, turn off the power. [It may cause a failure for
the Backup Controller.]
• Before attaching the Drive Motor, the ABD/Flow Sensor, the Temperature Sensor Cable, the Pressure Sensor Cable, the
Fast Clamp, the LAN cable and the CDI Communication Cable to the Controller, ensure the connectors and plugs are not
damaged. Similarly, before attaching the Drive Motor to the Backup Controller, ensure the connectors and plugs are not
damaged. If a connector pin is damaged or twisted, replace the system with a spare one and do not use it. [This may result
in decreased performance.]
• Since drug solution may cause a short circuit, ensure that the connecting sections (the Drive Motor, the ABD/Flow Sensor,
the Pressure Sensor Cable, the Temperature Sensor, the Fast Clamp, the LAN cable and the CDI Communication Cable)
are not wet when connecting plugs to connectors. If moisture is present, ensure that the power is turned off and the AC
power cable is removed from both this system and grounded AC power source, and then thoroughly wipe with a dry
soft cloth. [Since this system does not have a waterproof structure, drug solution and moisture may affect the electrical
components inside, and cause a malfunction.]
• This system should not be used in a place where vibration, dust, mist, corrosive gas, etc., occurs or in a place where the
system is sprayed with liquid. If liquid such as blood and drug solution is spilled on this system, thoroughly wipe with a dry
soft cloth. [This may result in decreased performance, and may cause a malfunction.]
• Do not touch the inside of the drive motor connector, the ABD/Flow sensor connector at the back of the Controller or the
drive motor connector at the front of the Backup Controller. Be sure to set up and operate within specified environmental
humidity levels. (Including attaching or detaching connector.) [It may cause static electricity resulting in failure or
malfunction.]
• When disposing of this system, dispose properly according to your local regulations.
(Controller/Backup Controller)
• Do not touch the AC power cable with wet hands and do not use a broken AC power cable. [Otherwise an electric shock or
short circuit may occur.]
• After circulation, make sure to rotate the motor speed knob to stop the Drive Motor and then turn off the power. [The Drive
Motor may fail.]
• Do not place objects near the AC power inlet and keep the AC power inlet away from walls. [The AC power cable cannot
be removed if other objects are in close proximity.]
• Do not place the Controller and the Backup Controller on a soft surface such as a bed. [The system may drop or fall. The
air inlet (with filter) at the bottom surface of the Controller and the Backup Controller may be closed.]
22
Summary of Contents for CAPIOX EBS Circuit with X coating
Page 1: ...CAPIOX Emergency Bypass System Instructions for use...
Page 4: ...CAPIOX EBS Circuit with X coating Instructions for use...
Page 8: ...Structure Structure...
Page 25: ...EBS and X coating are trademarks of TERUMO CORPORATION 2019 XX...
Page 26: ...CAPIOX EBS Cannula Kit with X coating Instructions for use...
Page 29: ...CAPIOX Centrifugal Pump Controller SP 200 Instructions for use...
Page 175: ...Memo 146...
Page 176: ...147 Memo...